6SHFLÀFDWLRQV
Warning: The technical specifications of the mask are provided for your healthcare
professional to determine if it is compatible with your CPAP or bi-level therapy device. If
used outside these specifications, or if used with incompatible devices, the mask may be
uncomfortable, the seal of the mask may not be effective, optimum therapy may not be
achieved, and leak, or variation in the rate of leak, may affect device function. The
pressure flow curve shown below is an approximation of expected performance.
Exact measurements may vary.
Pressure Flow Curve
0
10
20
30
40
50
60
70
80
90
0 5 10 15 20 25 30
Flow Rate (SLPM)
Pressure cm H
2
O
Amara Full Face Mask
Pressure Flow Curve
Resistance with Anti-Asphyxia Valve Closed to Atmosphere
Drop in Pressure at
50 SLPM 100 SLPM
All sizes 0.6 cm
H
2
O 1.5 cm H
2
O
Inspiratory and Expiratory Resistance with Anti-Asphyxia Valve Open to Atmosphere
Inspiratory Resistance 0.6 cm
H
2
O
Expiratory Resistance 0.7 cm H
2
O
Deadspace Volume
P 224 ml
S 231 ml
M 242 ml
L 253 ml
Sound Levels
A-weighted Sound Power Level 28 dBA
A-weighted Sound Pressure Level at 1 m 20 dBA
Disposal
Dispose of in accordance with local regulations.
Storage Conditions
Temperature: -4º to 140º F (-20º to +60º C)
Relative Humidity: 15% to 95% non-condensing