25
3.8 Sample Line Connection
As shown in the diagram, the sample gas pipeline is inserted into the
Quick Connect. Slowly open the needle valve until the flow reaches
about 1 SCFH (cubic feet per hour). Please allow time to let the air
samples purge the analyzer, you think you have fully cleared the
sample gas pipeline, wait for a minute or so to let the readings
stabilize. When you have completed the oxygen concentration
POA200
Quick
Connect
Tubing
Reducing valve
Sample
Gas