30
5 Maintenance & Replacement
1. Warranty: Within one year on the acquisition date, any material
and/or quality damage of the analyzer due to faulty workmanship or
defective parts is covered by warranty, but will not extend to damages
caused by careless and/or misuse of equipment.
2. Periodic calibration: Factory recommended calibration cycle is
every two weeks, but if the analyzer is used frequently and/or used in
harsh environments, shorter intervals are recommended.
3. The sensor replacement:
(i) If the sensor is found to be damaged, electrolyte may be leaking
from its’ chamber. Please replace the sensor.
(ii) If the sensor has been operational for 9 months. Please replace
sensor.
4. The sensor replacement steps
(1) Power off
(2) Have the replacement sensor ready.
(3) Open the sampling cover in the back of the analyzer. Incline the
analyzer to make sampling pool face down, shake the analyzer mildly,