Title LIFEPAK 15 Monitor Defibrillator Temperature Upgrade Page 2 of 34
Doc Type
Service Process Document
Doc ID
3308212
Doc Rev
C
Doc State
Released
State Date
05/21/2019
Purpose:
This instruction guides the Field Service Representative through the update process to add the temperature feature to a LIFEPAK
®
15
monitor/defibrillator.
Scope:
This upgrade is for 3206720-All (Version 1) and 3306272-All (Version 2) LIFEPAK 15 monitor/defibrillators Defibrillators that do not
previously have the Temperature feature installed.
These instructions are intended for authorized field service personnel who have reviewed the upgrade procedure and who are familiar with
the LIFEPAK 15 device.
All Version 1 (3206720-All) devices being upgraded to add the Temperature features must also have the V2 System PCB installed.
If the Invasive Pressure (IP) feature is installed, it must be removed before the device can be upgraded to add Temperature feature.
Temperature cannot be installed in conjunction with the IP feature.
When installing multiple upgrades into a LIFEPAK 15, the Temperature upgrade should be the last upgrade installed.
All upgrades that include V2 software have the voice recorder enabled. The Temperature setting will also be enabled. This will be
performed using the Combined Software Configurator Loader (CSCL) configuration steps at the end of this upgrade.
Training:
As a prerequisite, only personnel that have completed a LIFEPAK 15 monitor/defibrillator repair training class are allowed to perform this
update on a LIFEPAK 15 device. An individual will be considered fully trained when they meet the prerequisite, and read and understand
this document. Once this task is performed the individual has completed training. A documented training record will be retained.
Tools Required: (substitution of equivalent or better equipment allowed)
Computer – XP operating system
LIFEPAK 15 To PC Interconnect Cable (REF 11230-000005 or REF 11230-000020)
Torque Driver – Hand Operation, with an adjustable range of 5 – 12 in-lb.
Driver Bits – P1 (2”minimum shaft), P2 (5” minimum shaft), Torx, T-15 (2” minimum shaft)
Fixture Assembly, Capacitor Discharger 3012102
Tools called out in the PIP - TCP sections of LIFEPAK 15 PIP Test Procedure
All tools requiring calibration should be calibrated and within their calibration cycle dates.
• Torque Driver conversions – 4 lb-in = 64 in-oz = 45 cNm, 6.8 lb-in = 77 cNm, 10 lb-in = 113 cNm