Installation, use and maintenance
7
LONG RESET KEY KEYING
K900 is informed that the calibration procedure is finished.
Before performing this operation, make sure the indicated value is that required
8
NO OPERATION
At the end of the calculation, the new USER K FACTOR is shown for a few sec-
onds, aer which the restart cycle is repeated to finally achieve standby condition.
IMPORTANT: From now on, the indicated factor will become the calibration factor
used by the meter and will continue to remain such even aer a baery change
9
NO OPERATION
K900 stores the new calibration factor and is ready
for dispensing, applying the newly defined USER K FACTOR.
9 CONFIGURATION
K900 has a menu with which the user can configure the machine in accordance with his/her requirements.
To enter the configuration menu, proceed as follows:
1 wait until K900 is in Stand-by
2 press the CAL and RESET buttons at the same time and hold them down until the
word “Unit” and the previously-set unit of measurement appear on the display (Litre/
Litre in this example)
CALCAL
CAL
RESET
RESET
RESET