Technical Parameter
1) Model description: pM-KN01
Model pM-KN01
Rated power (VA) 550
Operation voltage
(V/Hz)
AC230±10% 50±1
Oxygen ow (L/mln) 0-5
Oxygen concentration
(%)
93%±3%
Outlet pressure (Mpa) 0.04—0.07
Alarm
Power failure low & high pressure ;
Sound level (dBA) ≤50
LCD display accumulating timing; present timing; timing;
Large LCD display
Switch times; pressure digital(accuracy:0.001MPa);
accumulating timing(range:0-10000hours);
present timing(accuracy:1 minute);
presetting timing(accuracy:1 minute)
Electrical category: Class II Type B
Product category: Class II a
Net Weight (Kg)
26
Fuse T6.3AL/250V
2) Nasal oxygen tube
Nasal oxygen tube is made up of intubatton, oxygen tube, inlet oxygen tube, movable gripper tube and nasal tube. And
it is made of medical PVC.There are two tips, over ear style. It must be exible hose, transparent and clean. There is no
impurities and no kink. When inputting the 50Kpa pressure the nasal oxygen tube must be non-disclosure continued
15S.The various components can withstand vertical stress of 20N in 15S without break. With Ethylene oxide sterilization,
since the date of sterilization, the nasal oxygen tube could be saved for three years.
3) Humidifier
The humidifier must be colorless and transparent or translucent. And it is marked with the maximum and minimum liquid
level instruction line. It should not be broken when it withstands the pressure of not less than 0.4Mpa.