30
Troubles and solution
No. Trouble Causes Solution
1 No operation after
power connected
1. No connection between circuit
of oxygen generator and power
2. Circuit of fuse protector broken
3. Capacitor of compressor broken
4. Compressor broken
1. Check out whether switch, plug, power line
in good connection.
2. Replace the fuse protector and nd the
cause
3. Replace start capacitor
4. Have the compressor replaced
2 No oxygen out or
tiny outtake ow
1. Folded inside oxygen tube, no
smooth outtake
2. Filter clogged, no smooth intake
3. The cover of dampen bottle
leaking
1. Connect the oxygen tube again
2. Clean the lter
3. Take off the cover, screw well the cover,
block the outtake by thumb after turning
on, and there will some sound from the
humidier after 5 second around (the safety
valve of humidier turns on)