Set-Up Sub-Menus
Press the
key to access the Temperature Controller’s set-up sub-menus. The
or
keys are used
to change the current setting / value in the sub-menu.
Sub-Menu Selection / Range
To Change
Current Value
Temperature unit °C or °F
Press
for °C
Press
for °F
Offset Calibration -3.0°C to +3.0°C Press
or
Low Limit -35° to +10°C / -30° to +50°F Press
or
High Limit +40° to +140°C / +100 to +285°F Press
or
Auto Cool
1° to 135°C / 34° to 275°F (15L and larger reservoirs)
1° to 70°C / 34° to 158°F (7L reservoirs)
Press
or
To accept a value in a sub-menu, press
or allow the display to ‘time out’ (approximately 10 seconds).
Adjusting the Temperature Set Point
This is the temperature at which the fluid in your Circulating Bath will be maintained. It may be set to one-
tenth of a degree over a range of -30.0° to +135.0°C / -20.0° to +275.0°F. The factory default set point is
+20.0°C / +68.0°F.
To Change: Press the
or
key. The word “SET” will begin flashing. Continue holding
or repeatedly press
or
until the desired set point temperature is displayed.
To Accept: Press
or allow the LCD to return to the main operational display
(approximately 10 seconds).
Fl
ashes when set
point is being
changed
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PolyScience MX07R-20 and is the answer not in the manual?

PolyScience MX07R-20 Specifications

General IconGeneral
BrandPolyScience
ModelMX07R-20
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals