User Guide
DuraChill Air- and Water-Cooled 1.5 HP Chillers
110-392
27 June 2016

Table of Contents

Questions and Answers

  • R
    Ryan DavisAug 13, 2025
    Why is my PolyScience DuraChill 58 not cooling enough?
    • D
      david42Aug 13, 2025
      If your PolyScience Laboratory Equipment chiller isn't cooling properly, several factors could be the cause. Check for dust buildup on the air filter or condenser and clean them if necessary. Ensure that the air ventilation screens are not blocked. Verify that the heat load doesn't exceed the chiller's capacity. The ambient air temperature might be too high, so try to decrease it. Also, check for low or high line voltage and correct it as needed. If the temperature readings of the compressor upper, evaporator inlet and evaporator outlet temperature sensor is -50°C, the sensor needs to be replaced.
  • L
    Lisa KellyAug 17, 2025
    What to do if there is no fluid circulation in my PolyScience DuraChill 58?
    • T
      Taylor GoodmanAug 17, 2025
      If you're experiencing no fluid circulation in your PolyScience Laboratory Equipment, there are a few things to check: * Make sure there is enough fluid in the reservoir and add more if needed. * Look for any blockages in the circulating system and remove them. * If the pump isn't operating, it may need to be replaced.
  • A
    Alexander JacksonAug 20, 2025
    How to improve insufficient circulation in PolyScience DuraChill 58 Laboratory Equipment?
    • M
      Michael JonesAug 20, 2025
      If you have insufficient circulation in your PolyScience Laboratory Equipment, consider the following: * The fluid viscosity might be too high; replace it with a lower viscosity fluid. * The external tubing diameter could be too small; replace it with larger diameter tubing. * Check for any restrictions in the fluid lines and correct them. * Also, check for low line voltage and correct it as required.
  • R
    Rebekah UnderwoodAug 25, 2025
    Why my PolyScience DuraChill 58 doesn't run and the display is blank?
    • S
      Sydney DavisAug 26, 2025
      If your PolyScience Laboratory Equipment unit does not run and the digital displays are blank, check that the electrical cord or wiring is securely connected to a working electrical source. Also, make sure the Power Switch / Circuit Breaker on the rear of the unit is turned ON.
  • J
    Jill SmithAug 28, 2025
    What to do if PolyScience DuraChill 58 doesn't run but shows decimal points?
    • J
      Jessica LawsonAug 28, 2025
      If your PolyScience Laboratory Equipment unit does not run and three decimal points appear on the Temperature Display and two decimal points on the Pressure/Flow Rate Display, the unit might be in Standby mode. Try pressing the Power Button on the front panel to turn it on.
  • J
    Jonathan ClarkSep 1, 2025
    Why is my PolyScience DuraChill 58 Laboratory Equipment not heating?
    • I
      idawsonSep 1, 2025
      If your PolyScience Laboratory Equipment does not heat, it might be because the OTP has tripped due to insufficient fluid in the reservoir. Add fluid to the reservoir and then reset the OTP.
  • C
    Cynthia LeachSep 5, 2025
    What does fault code 10 mean on my PolyScience DuraChill 58 Laboratory Equipment?
    • T
      Tracy StewartSep 5, 2025
      If fault code 10 appears on the display of your PolyScience Laboratory Equipment, it could be due to extreme electrical line interference. Try plugging the unit into another power source. If the problem continues, the triac may have failed, and you should replace the triac or triac driver as required.
  • D
    Daniel ElliottSep 8, 2025
    What does fault code 12 mean on PolyScience DuraChill 58 Laboratory Equipment?
    • A
      Andrew PotterSep 8, 2025
      If you see fault code 12 on your PolyScience Laboratory Equipment display, it could indicate a loose external probe connection. Check and correct it as needed. Alternatively, the external temperature probe might be faulty and need replacement. Note that the chiller can still be operated using the internal probe until the issue is resolved.

Summary

General Safety Information

Safety Recommendations

User responsibility to read and understand all safety precautions before operating the Chiller.

General Safety Warnings

Crucial warnings regarding operation, service, and handling of the Chiller to prevent hazards.

Controls and Components

Control Panel

Overview of the control panel layout and its interactive components.

Installation and Startup

Process Fluid Connections

Details on making process fluid connections to the Chiller's inlet and outlet.

Facility Water Connections

Instructions for connecting facility water supply and return lines for water-cooled models.

Process Coolant: Suitable Fluids

Lists approved and prohibited fluids for use as coolant in the Chiller.

Electrical Power Requirements

Guidelines and warnings for connecting electrical power to the Chiller.

Optional Signal Inputs;Outputs

Details on connecting optional communication and control interfaces.

Startup Procedure

Step-by-step instructions for starting up the Chiller.

Starting Process Fluid Flow

Steps to initiate process fluid circulation after startup.

Normal Operation

Displaying and Adjusting Set Point

Procedure for viewing and changing the desired temperature set point.

Operational Parameters and Limits

High Temperature Limit (HL)

Sets the maximum allowable set point temperature with alarms.

Low Temperature Limit (LL)

Sets the minimum allowable set point temperature with alarms.

Maximum Fluid Pressure (FP)

Sets the maximum allowable fluid pressure with alarms and shutdown.

Minimum Flow Rate (FL)

Sets the minimum allowable fluid flow rate with alarms and shutdown.

Calibration Offset (C1;C2)

Adjusting displayed temperature to match a traceable standard.

Display, Alarm and Error Messages

Message Code Descriptions

Explanation of message codes, their descriptions, and required actions.

Specific Fault Codes

Adjusting High Pressure Bypass Setting

Over-Temperature Protection

Routine Maintenance

Temperature Calibration

Troubleshooting

Problem Analysis and Corrective Actions

Lists common problems, possible causes, and corrective actions.

Diagnostic Mode

Technical Information

Performance Specifications

Technical performance data for Air-Cooled and Water-Cooled models.

Electrical Specifications

Electrical requirements and ratings for various Chiller models.

RS-232; RS-485 Communications Protocol

Details on serial communication protocols and commands.

Service and Technical Support

Appendix

Electrical Diagram (208-230 V;1 PH;50-60 Hz No Heater)

Wiring diagram for 208-230V, 1PH, 50-60Hz models without heater option.

Electrical Diagram (208-230 V;1 PH;50-60 Hz With Heater)

Wiring diagram for 208-230V, 1PH, 50-60Hz models with heater option.

Electrical Diagram (208-230 V;3 PH;50-60 Hz No Heater)

Wiring diagram for 208-230V, 3PH, 50-60Hz models without heater option.

Electrical Diagram (208-230 V;3 PH;50-60 Hz With Heater)

Wiring diagram for 208-230V, 3PH, 50-60Hz models with heater option.

PolyScience DuraChill 58 Series Specifications

General IconGeneral
Temperature Range-20°C to +40°C
RefrigerantR404A
ControllerDigital PID
Safety FeaturesOver-temperature protection, low liquid level protection
CommunicationRS232