Installation and Startup
Site Requirements
Ambient Temperature and Relative Humidity
The Chiller is designed for indoor installation in ambient temperatures between 16° and 35°C (60° and 95°F);
relative humidity should not exceed 80% (non-condensing).
Location
Install the Chiller on a strong, level surface as close to possible to the process requiring cooling. It should not be
installed closer than 4 feet (1.4 meters) to a heat-generating source. If possible, the unit should be located near a
suitable drain to prevent flooding in the event of leaks. Do not place it where corrosive fumes, excessive moisture,
excessive dust, or high room temperatures are present.
Position the Chiller as close as possible to the power distribution panel. Avoid voltage drops by using a properly
grounded power source wired to meet electrical data plate requirements. The use of an extension cord is not
recommended.
NOTE: The unit may be located at a level below that of the equipment being cooled. As long as the process
remains closed, overflow will not occur when adding cooling fluid to the unit's reservoir.
Clearance
Allow adequate clearance on the front, sides, and rear of the Chiller for access to connections and components.
The front and rear vents of the Chiller must be a minimum of 24 inches (61 cm) away from walls or vertical
surfaces so air flow is not restricted.