When Not to Use the Posey Bed (Cont.)
Zippers and zipper pins are not
completely seated in their boxes and/
or not all of the teeth are engaged.
Zippers produce gaps and/or openings
when pressure is applied to the entire
length of the zipper.
Zippers do not close securely along the
entire length. Gaps or openings in the
zipper may allow the patient to open
the zipper and fall out of bed. Always
apply pressure along the entire length
of the closed zipper to make sure there
are no gaps or openings.
Zipper pull tabs are open/damaged and
will not close the zippers.
Crown (Tab) Holder
Body
Pull-Tab
Undamaged pull tab
Opened Up
Damaged pull tab