EasyManua.ls Logo

Power Plate pro6 - Electromagnetic Immunity Guidance

Power Plate pro6
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The
pro6
+
is intended for use in the electromagnetic environment specied below. The customer or user of the
pro6
+
should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or
discharge (ESD) ±8 kV air ±8 kV air ceramic tile. If oor are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be at least 30%.
Electrical fast ±2 kV for power ±2kV for power Main power quality should be that of
transient/burst supply lines supply lines a typical commercial or hospital IEC
61000-4-4 environment.
Surge ± 1 kV line(s) t ±1 kV differential Main power quality should be that of
IEC 61000-4-5 o line(s) mode a typical commercial or hospital
environment.
Voltage dips, short <5% U
T
<5% U
T
Main power quality should be that
interruptions and (>95% dip in U
T
) (>95% dip in U
T
) of a typical commercial or hospital
voltage variations for 0.5 cycle for 0.5 cycle environment. If the user of the
pro6
+
on power supply requires continued operation during
input lines 40% U
T
40% U
T
power main interruptions, it is IEC 61000-
4-11 (60% dip in U
T
) (60% dip in U
T
) recommended the
pro6
+
be powered
for 5 cycles for 5 cycles from an uninterruptible power supply or a
batter y.
70% U
T
70% U
T
(30% dip in U
T
) (30% dip in U
T
)
for 25 cycles for 25 cycles
<5% U
T
<5% U
T
(>95% dip in U
T
) (>95% dip in U
T
)
for 5 sec for 5 sec
Power frequency
3A/m
3A/m
Power frequency magnetic elds should
(50Hz) magnetic be at levels characteristic of a typical
61000-4-8
location in a typical commercial or
hospital environment.
NOTE U
T
is the a.c. main voltage prior to application of the test level.
11
of 22

Other manuals for Power Plate pro6

Related product manuals