! Electromagnetic Compatibility (EMC)
Precautions
The pro6
+
model is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Both models have been tested against,
and have passed the applicable
requirements of relevant electromedical
standards, including EN 60601-1-2:2007.
The pro6
+
model emits electromagnetic
energy to perform its intended function.
Nearby electronic equipment may be
affected by this emission. Similarly
the pro6
+
model may be affected by
electromagnetic emissions from other
equipment in the vicinity.
The pro6
+
model is intended for use
in the electromagnetic environment
specied below. The customer or user of
the pro6
+
model should ensure that it is
used in such an environment:
Floors should be wood, concrete or
ceramic tile. If the oor is covered with
synthetic material, the relative humidity in
the room should be at least 30%.
Power quality should be that of a typical
domestic, commercial or hospital
environment as appropriate.
The pro6
+
model should not be used
during power interruptions.
Power frequency magnetic elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
The pro6
+
model is intended for use in an
electronic environment in which radiated
radio-frequency (RF) disturbances
are controlled.
The customer or user of the pro6
+
model can help prevent electromagnetic
interference by maintaining a minimum
distance between portable and mobile
RF communications equipment
(transmitters) and the pro6
+
model as
recommended in this user manual,
according to the maximum output power
of the communications equipment:
The adequate separation distance
to provide base immunity to RF
disturbances is 1.0m.
USB Port - USB port supplies 5Volt DC
current. This port is only used for soft-
ware updates and uploading personal
pictures to user proles. The operator
must take care to not touch the USB port
and a patient simultaneously.
LAN port - the pro6
+
has included a
LAN port on the lower back side of
the machine. This LAN port is used to
connect a LAN cable to access the
internet.The operator must take care to
not touch the LAN port and a patient
simultaneously.
Phone jack adapter- On the control
panel of the pro6
+
there is a phone
jack adapter for sound playback. The
operator must take care to not touch
the phone jack adapter and a patient
simultaneously.
Retain this user manual for your future
reference.