Guidance and manufacture’s declaration – electromagnetic emissions-
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The Pro7 is intended for use in the electromagnetic environment specied below. The customer or the user
of the Pro7 should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
Group 1 The Pro7 use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interfe rence in nearby electronic
equipment.
RF emission CISPR 11 Class B Class A with IEC61000-3-2
Complies with IEC61000-3-3
Harmonic emissions Class A The Pro7 is suitable for use in all establishments,
IEC 61000-3-2 including domestic establishments and those directly
connected to the public low-voltage power supply
Voltage uctuations/ Comply network that supplies buildings used for domestic
icker emissions purposes.
IEC 61000-3-3
Wi-Fi Specications Network Standard: IEEE 802.11b/g/n
Band: 2.400-2.4835GHz
Cover Range: 20meter
Wi-Fi Power: 10dBm
EMC Declaration
Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The Pro7 is intended for use in the electromagnetic environment specied below. The customer or the user of
Pro7 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or
discharge (ESD) ±8 kV air ±8 kV air ceramic tile. If oor are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be at least 30%.
Electrical fast ±2 kV for power ±2kV for power Mains power quality should be that of
transient/burst supply lines supply lines a typical commercial or hospital IEC
61000-4-4 environment.
Surge ± 1 kV line(s) t ±1 kV dierential Mains power quality should be that of
IEC 61000-4-5 o line(s) mode a typical commercial or hospital
environment.
Voltage dips, short <5% U
T
<5% U
T
Mains power quality should be that
interruptions and (>95% dip in U
T
) (>95% dip in U
T
) of a typical commercial or hospital
voltage variations for 0.5 cycle for 0.5 cycle environment. If the user of the Pro7
on power supply requires continued operation during
input lines 40% U
T
40% U
T
power mains interruptions, it is IEC
61000-4-11 (60% dip in U
T
) (60% dip in U
T
) recommended that the Pro7 be powered
for 5 cycles for 5 cycles from an uninterruptible power supply or a
battery.
70% U
T
70% U
T
(30% dip in U
T
) (30% dip in U
T
)
for 25 cycles for 25 cycles
<5% U
T
<5% U
T
(>95% dip in U
T
) (>95% dip in U
T
)
for 5 sec for 5 sec
Power frequency
3A/m
3A/m
Power frequency magnetic elds should
(50Hz) magnetic be at levels characteristic of a typical
61000-4-8
location in a typical commercial or
hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.