31
Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and this Device:
This device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the device can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and this device as recommended below,
according to the maximum output power of the communications equipment.
Rated Maximum Power
Output of Transmitter
(W)
Separation Distance According to Frequency of Transmitter (M)
150 kHz to 80 MHz
d=1.2√P
80 MHz to 800 MHz
d=1.2√P
800 MHz to 2.5 GHz
d=2.3√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
The guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Live Active Five POC is intended for use in the electromagnetic environment
specified below. The user of the device should assure that it is used in such an
environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1 The device uses RF energy only for its
internal function. Therefore its RF emissions
are very low and not likely to cause any
interference in nearby equipment.
RF emissions
CISPR 11
Class B The device is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complies