3.2 Warning Lamp
All X-RAD Biological Irradiators feature an “X-Ray On” warning lamp to give the users and bystanders
a visual notification that X-Rays are being produced.
Note-If the warning lamp fails, X-Rays will shut off, a “Warning Lamp” error will be displayed to the
User Interface and X-Rays will be disabled until the Warning Lamp is replaced.
3.3 Baffle
The baffle is a lead shielding around the port hole(s) of each X-RAD Biological Irradiator to stop/limit
irradiation leakage outside of the cabinet. Cabinets containing baffle safety interlocks feature
redundant interlocks ensuring the baffle cannot be open and X-Rays produced. Some X-RAD models
feature dedicated user accessible baffles to allow the user to run wires, cables, hoses and equipment
into the cabinet without any radiation leakage. If an X-RAD model does not have a dedicated user
accessible baffle, the user can run wires, cable, hoses and equipment through the main baffle.