31
Section 9 Display Messages
(
(
c
c
o
o
n
n
t
t
i
i
n
n
u
u
e
e
d
d
)
)
Display: What It Means: What to Do:
Standard Strip not
inserted fully into
Test Strip Platform
or Meter was
moved during test.
Meter is cold. May
not give correct
result.
Meter is hot. May
not give correct
result.
Turn Meter off.
Repeat test
(see
Sec. 3)
without
moving Strip or
Meter.
Move Meter and
Strips to warmer
place (59º-98ºF or
15º-37ºC). Allow
Meter and Strips
to adjust to
temperature. Run
Standard Strip test
before using Meter
(see Sec. 3).
Move Meter and
Strips to cooler
place (59º-98ºF or
15º-37ºC). Allow
Meter and Strips
to adjust to
temperature. Run
Standard Strip test
before using Meter
(see Sec. 3).