DEFLATION MARK
INFLATION MARK
SYSTOLIC mmHg
DIASTOLIC mmHg
PULSE RATE BPM
PULSE RATE MARK
IRREGULAR
PULSE RHYTHM
MEMORY
BANK
NUMBER
MEASURING METHOD
An example of irregular pulse
ryhthm compared to regular pulse
rhythm
1.
2.
3.
Press START button.
The cuff is automatically inflated to the applied pressure.
Pressure (displayed value) starts to decrease and pulse is indicated
by heart symbol.
When the measurement is complete, air is automatically released
from the cuff.
Maximum and minimum blood pressures and pulse
rate are displayed.
Heart symbol flashes when pulse rhythm detected
during measurement was irregular.
Press either one of memory buttons, M1 or M2,
and the result is stored in the selected bank.
The selected memory bank number is shown on the display. The
result is stored in the bank selected when the unit is turned off.
Irregular pulse rhythm indication will not be stored, only blood
pressure and pulse rate are stored. When a measurement
resulted in an error, it will not be stored.
Press START button to turn off the power.
If you forget to turn off the unit, it will automatically turn off after
3 minutes.
W
ait 2 -3 minutes between
repeated measurements.
Automatic Repressurization
If the pressure applied is judged insufficient in the early phase of measurement
or if movement of the hand or wrist occurs during measurement, the unit
will inflate again to a level about 40 mmHg higher.
Automatic Repressurization
is repeated until a measurement is made. However, this does not indicate
a problem.
To manually control pressurization, hold down START button immediately
after the start of inflation and release it when the pressure reaches the level
50 mmHg higher than the expected maximum systolic pressure. If the
pressure value has exceeded 180 mmHg, the inflation will stop when the
key is released. Pressure can be increased to approximately 300 mmHg.
regular pulse rhythm
Irregular pulse rhythm indication
Heart symbol flashing in measurement result display indicates
irregular pulse rhythm.
Irregular pulse rhythm can result from
body movement during measurement or can be caused by
arrhythmias. Although constant appearance of the indication under
quiet measurements may suggest arrhythmias, do not make any
judgment on your own before your doctor's consultation.
To stop measurement
Press START button and
the unit stops inflation,
discharges air rapidly,
and then turns off.
-8-