EasyManua.ls Logo

Princeton Instruments PIXIS System - Figure 27. Binning and Array Orientation

Princeton Instruments PIXIS System
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Operation 63
63
Figure 27. Binning and Array Orientation
Note: You can easily switch between these orientations by rotating the camera 90° and
changing the binning parameters in the application software.
Output Amplifier {Quality} Selection
The output amplifier amplifies the collected charge from the output node and outputs it as
electrons/count. Although Figure 24 shows an array with dual output nodes and
amplifiers (one set at each end of the shift register), some PIXIS systems are available
with a single output node and amplifier. If your system has dual output amplifiers, you
can choose the output amplifier to be used (High Capacity or Low Noise) via the
Acquisition|Experiment Setup…|ADC tab {Quality on the Analog to Digital
Conversion expander}:
High Capacity amplifier: Provides a spectrometric well capacity that is
approximately 3 times the well capacity for the Low Noise amplifier selection.
High Capacity is suitable when you have intense light signals or signals with
high dynamic range.
Low Noise amplifier: Provides the highest sensitivity performance and is
suitable when you have weak signals.
Note: The choice of output amplifier {Quality} and controller gain {Analog Gain}
setting should be considered together for the best signal capture. Examples of the
interaction of output amplifier and controller gain selections are shown in Table 2.

Table of Contents