Special Instructions
Notes:
1. Failure to operate, store and use the Quantus™ Fluorometer as directed
will invalidate the warranty.
2. Direct plug-in power supply acts as a disconnect. To disconnect power,
pull power supply from the outlet.
Changes or modifications to this unit not expressly approved by the
party responsible for compliance could void the user’s authority to
operate the equipment.
Do not use this device in proximity to sources of strong electromagnetic
radiation (e.g., unshielded intentional RF sources), because they may
interfere with the proper operation.
Do not use this instrument for anything other than its intended use.
If the instrument is used in a manner other than that specified by
Promega, the protection provided by the equipment may be impaired.
Do not disassemble the unit.
The Quantus™ Fluorometer accepts 0.5ml PCR tubes.
Do not force larger tubes into the tube holder.
Use caution around solvents because they may corrode the plastic case of
the Quantus™ Fluorometer.
Do not expose the Quantus™ Fluorometer to temperatures outside the
specified range since damage to the unit may occur.
Avoid bubbles in your sample because bubbles may negatively affect the
fluorescence reading.
Always disconnect the power before cleaning or performing routine
maintenance.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA. Part# TM396
Revised 1/20 Page 5
Precautions (continued)
Always disconnect the AC
Adapter from the power outlet
when not in use.
Toujours enlever l'adaptateur AC
de la prise électrique lorsqu'il n'est
pas utilisé.