1.E. Environmental Requirements
Operations, Shipping and Storage Conditions
• The instrument is intended for indoor use only.
• Install on a sturdy, dust-free, level surface.
• Do not install in a location where there is large temperature variability or
high humidity.
• Position the instrument so that it is easy to unplug from the power
source.
• Do not place the power supply in areas where it may come in contact
with liquid.
• Do not place next to heat sources.
• Do not use near flammable gases or liquids.
• For optimal performance, we recommend equilibrating the instrument to
operating temperature before use.
2. Quantus™ Fluorometer Inspection and Setup
Upon receiving your Quantus™ Fluorometer, inspect everything carefully and
make sure all items are present and undamaged.
2.A. Setting Up the Quantus™ Fluorometer
1. Position the Quantus™ Fluorometer on a flat surface.
2. Select the appropriate outlet adapter for your wall outlet from the Outlet
Adapter Kit provided.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Part# TM396 Printed in USA.
Page 6 Revised 1/20
Operating Conditions 15–30°C, up to 75% humidity
Operating Altitude Up to 2,000 meters
Shipping and Storage Conditions
5–40°C under noncondensing
conditions, up to 75% humidity
Maximum Relative Humidity
Up to 80% for temperatures up to
31°C, decreasing linearly up to 50%
relative humidity at 40°C
MAINS Power Supply
Supply voltage fluctuation up to ±10%
of the nominal voltage
Transient Overvoltage
Up to the level of Overvoltage
Category II
Temporary Overvoltage
Occurring on the MAINS power
supply
Pollution Degree Pollution Degree 2
TM396.0120_EIVD_TM.qxd 1/27/2020 2:50 PM Page 6