TEKTINO AC-200-dual USER MANUAL V2.30
14
/14
During recovery,
vacuum pump is
pressurized. After
period too much oil
in vacuum pump
The contact between solenoid valve
#8 and valve base is not well sealed.
Remove solenoid #8 from valve
base, clean the solenoid valve and
valve base.
During vacuum,
there is suction in old
oil bottle.
The contact between solenoid valve
#2 and valve base is not well sealed.
Remove solenoid #2 from valve base, clean
the solenoid valve and valve base.
Remarks: Regular maintenance by specialized technicians is highly recommended.