Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Service Manual - Cleaning
38 D596.131.00/05 en
Clean the product simply by wiping it down with
a dry or slightly moistened cloth. Use water or a
mild soap solution.
Alternatively, you can use a common disinfection
agent for medical instruments which is specified
in the list of disinfectants of your national hygiene
institute (e.g., VAH in Germany). If your country
does not provide such a list, we recommend either
a solution on an aldehyde base or a solution with
a quaternary ammonium compound.
Detectors and Connection Cables
•
Disconnect the detectors from the measuring
device (always pull on the connector, not on
the cable!).
•
Only wipe the detectors and the connection
cables down. Do not immerse the detectors or
connection cables in liquid. Liquids must not
be allowed to penetrate. Do not spray anything
onto the detectors and connectors. All con-
tacts must remain clean and dry.
Connectors
NOTICE
To avoid equipment damage, do not immerse
the connectors in liquids. Liquids must not be
allowed to enter the connectors.
The connectors must always be clean. Use a soft
brush to remove dust (do not blow onto the con-
nector!).
Non-use
When not in use, the product should be stored in
its case for protection.
The connector of the detector should be protected
against humidity and dust with the supplied cap.
The chamber volume must be protected against
mechanical damage by means of the build-up
cap.