Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual
D596.131.00/05 en 29
4 Electromagnetic Compatibility according to IEC 60601-1-2
The following statements apply to the Farmer
chambers types 30010, 30011, 30012, and
30013.
The Farmer chambers will be referred to as prod-
uct in these statements.
WARNING
Magnetic and electrical fields are capable of
interfering with the proper performance of the
product.
Product Failure!
Do not use cables longer than specified. This
might impair the electromagnetic compatibility
characteristics of the product.
WARNING
Magnetic and electrical fields are capable of
interfering with the proper performance of the
product.
Product Failure!
Do not use the product in the immediate vicin-
ity or placed on top of or below other equip-
ment. However, if the application requires an
arrangement of devices as described above,
the product should be monitored in order to
ensure its proper functioning in the specific
arrangement.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the
user of the product should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
IEC/CISPR 11
Group 1 The product uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby elec-
tronic equipment.
RF emissions
IEC/CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
not applicable
Voltage fluctuations/flicker
emissions IEC 61000-3-3
not applicable
The product is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply net-
work that supplies buildings used for domestic pur-
poses.
1)
1)
The product fulfills the requirements of IEC/CISPR 11 Class B. However, it is not intended for use in domestic
establishments.