Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Service Manual - Disposal of the Product
40 D596.131.00/05 en
Calibration
The exact calibration factor and supplementary
data are included in the calibration certificate.
For use at high energies (above 1.3 MeV cor-
rection factors according to the relevant national
or international dosimetry protocols must be
observed (e.g., IAEA TRS 277 or TRS 398).
The product should be recalibrated at a dose rate
typical for the application. Calibrations should be
done at
60
Co.
It is recommended to have the product calibrated
every 2 years at PTW-Freiburg or by specially
trained staff on site (when using a radioactive
check device and performing regular check mea-
surements: 6 years).
8 Disposal of the Product
The typical lifetime of the product is approx.
10 years, if used normally.
At the end of the product lifetime, the components
of the product must be disposed of in compliance
with the applicable waste control regulations. The
different materials must be separated and recy-
cled as appropriate.
The electronic components must be recycled
according to the local regulations.
The product does not contain hazardous materi-
als.