Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual
30 D596.131.00/05 en
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the
user of the product should assure that it is used in such an environment.
Immunity
tests
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV
air
not applicable Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humid-
ity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
for power supply
lines
± 1 kV
for input/output lines
not applicable
± 1 kV
for input/output lines
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Surges
IEC 61000-4-5
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
not applicable Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5 % U
T
(> 95 % dip in U
T
) for
½ cycle
40 % U
T
(60 % dip in U
T
) for
5 cycles
70 % U
T
(30 % dip in U
T
) for
25 cycles
< 5 % U
T
(> 95 % dip in U
T
) for
5 s
not applicable Mains power quality should be that of
a typical commercial or hospital envi-
ronment. If the user of the product
requires continued operation during
power mains interruptions, it is rec-
ommended that the product be pow-
ered from an uninterruptible power
supply or a battery.