Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Technical Manual
D596.131.00/05 en 31
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Immunity
tests
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.