Farmer Chamber
Ionization Chamber
Type 30010, 30011, 30012, 30013
Safety Information
D596.131.00/05 en 9
DANGER
Electricity is a source of risk, particularly when
the product is operated inappropriately.
Shock Hazard!
Exclusion of operation in the patient envi-
ronment:
Do not operate the product nor any periph-
eral devices in the patient environment (see
Figure 3).
Exclusion of operation with patient contact:
The product is not intended for use in direct
contact with the patient. Ensure that neither the
product nor any peripheral devices come into
contact with the patient.
DANGER
Electricity is a source of risk, particularly when
the product is operated inappropriately.
Shock Hazard!
Use of peripheral devices:
Connect only peripheral devices (PC, printer)
that meet the requirements of the standard
IEC 60950-1 (UL 60950-1) or the standard
IEC 61010-1.
Connect devices to other devices or to parts
of systems only if it has been ascertained that
this connection does not impair the safety of the
patient, the operator, or the environment.
If the device specifications do not contain infor-
mation regarding the connection of the device
to other equipment, consult, for example, the
manufacturers of the other equipment or other
informed experts to verify that the connection
does not impair the safety of the patient, oper-
ator, or environment. Always observe the stan-
dard IEC 61010-1.