9
8
Secure the cable by turning clockwise the locking ring of the plug (16).
Connect an external power supply (AC/DC) or car adapter to the jack 
(17) of the cable. 
Note: central pin of the power supply that you connect to the “power” jack 
of the sight, must have marking “+”.  
The power supply may have marking             .
Connection of an external power supply (icon       on the status bar) 
automatically cuts off power supply from batteries. 
External power supply DOES NOT charge the batteries in the sight!  
Attention! We  suggest that you  use battery  packs  EPS3  or EPS5 
ensuring from 9 to 20 hours of continuous operation.
OPERATION
8
WARNING! Do not point the objective lens of the unit at intensive 
sources of light such units emitting laser radiation or the sun. This 
may disable unit’s electronic components. The warranty does not 
cover damage caused by improper operation.
Switching on and calibration
Press “ON” (6) to switch on the unit. Green LED indicator (12) will light 
up. In case of low battery the LED will turn red and battery icon         
will start flashing.  
Do  the  calibration  of  image.  Calibration  levels  background 
temperature of the microbolometer and eliminates image flaws.  
There are  three calibration modes: manual (M), semiautomatic  (H) 
and automatic (А).
Semiautomatic calibration mode
Turn on the unit, open the lens cap. 
Press and hold the controller (2) for two seconds to enter the menu.
Rotate the controller to select option Cal. Press the controller.
Rotate the controller to select mode H. Press the controller to confirm. 
Press the CAL (5) button to calibrate. The image will freeze for 1-2 
seconds.
Calibration is completed.
Automatic calibration mode 
With the automatic calibration mode the thermal imager calibrates by 
itself  according  to  the  software  algorythm.  The  detector 
(microbolometer) is closed with the shutter automatically. User assisted 
calibration with the CAL (5) button is allowed in this mode.   
Turn on the unit, open the lens cap. 
Press and hold the controller (2) for two seconds to enter the menu.
Rotate the controller to select option Cal. Press the controller.
Rotate the controller to select mode A. Press the controller to confirm. 
At the moment of the automatic calibration the image will freeze for
1-2 seconds. 
Calibration is completed.
Focusing and image adjustment
Open the lens cap (1).
Adjust  sharp  image  of  the  display  icons  by  turning  the  eyepiece 
diopter adjustment ring (13). 
To  adjust  display  brightness,  rotate  the  controller  (2).  Respective 
brightness level (from 0 to 20) appears next to brightness icon
in the upper right portion of the display.    
To adjust  display contrast, press briefly  the controller (2) (contrast 
icon        appears) and rotate it. Respective contrast level (from 0 to 
20) appears next to contrast icon in the upper right portion of the 
display.     
Point the unit at  a  warm  object located at a  certain  distance,  100 
meters, for example.
Achieve a sharp image by turning the lens focus ring (7).  
After  this  adjustment  no  further  dioptre  adjustment  should  be 
required, regardless of distance or other factors. Adjust image quality 
only with the lens focus ring (7).
Manual calibration mode
The lens cap (1) should be closed (in LQ/XQ19 models a diaphragm 
(15) inside the objective lens plays  the  role  of  the cap. Rotate the 
wheel  (14)  clockwise  to  open  the  diaphragm,  counterclockwise  to 
close it). Turn on the unit, press and hold the controller (2) for two 
seconds to enter the menu.
Rotate the controller to select option Cal. Press the controller.
Rotate the controller to select mode М. Press the controller to confirm. 
To exit the menu, press and hold the controller (2) for two seconds or 
wait 10 seconds for automatic exit.
Press the CAL (5) button to calibrate. The image will freeze for 1-2 
seconds. Then open the lens cap. Calibration is completed. 
In case you see image flaws (such as frozen image, vertical stripes 
etc.) re-calibrate the unit.