51
Electromagnetic Immunity
The Milk Genie is intended for use in the electromagnetic environment
specified below. The customer or the user of the pump should ensure that
it is used in such an environment.
Immunity
Test
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 ㎸ Contact
±8 ㎸ air
±6 ㎸ Contact
±8 ㎸ air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrostatic
Discharge (ESD)
FDA Home Use
Guidance
±8 ㎸ Contact
±15 ㎸ air
±8 ㎸ Contact
±15 ㎸ air
Electrical Fast
Transient/Burst
IEC61000-4-4
±2 ㎸ for
power supply
lines
±1 ㎸ for
input/output
lines
±2 ㎸ for
power supply
lines
±1 ㎸ for
input/output
lines
Main power quality should
be that of a typical
commercial or domestic
environment.
Surge IEC61000-
4-5
±1 ㎸ line(s) to
line(s)
±2 ㎸ line(s) to
earth
±1 ㎸ line(s) to
line(s)
±2 ㎸ line(s) to
earth

Other manuals for Pumpables Milk Genie