EasyManuals Logo

Qsonica SONICATOR S-4000 User Manual

Default Icon
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
23
The following Option Screen can be accessed from either Menu Screen.
The Option menu enables the selection of Temperature units, Footswitch operational
modes and Display contrast.
Option Screen
Temperature Units – A temperature probe (# TP4000) is required for use of this option.
The Temperature Probe option can be selected if you wish to monitor the temperature of the
sample being processed.
The temperature probe must be plugged into the back of the Sonicator and the probe tip
must be in the liquid sample. If the initial sample temperature is above the set point,
sonication will not turn on. During sonication, if the temperature reaches the Shutdown
value, the unit will pause sonication. The unit will remain paused until the temperature drops
below the pre-set temperature limit, at which time sonication can be manually resumed.
Footswitch Operation – A footswitch (# FS-3) is required for use of this option.
After setting up a program, the footswitch can be used to remotely activate the unit.
Display Contrast – The left arrow will lighten the screen, the right arrow will darken the
screen.
Exit – The Exit button will save changes and return to the previous screen.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Qsonica SONICATOR S-4000 and is the answer not in the manual?

Qsonica SONICATOR S-4000 Specifications

General IconGeneral
BrandQsonica
ModelSONICATOR S-4000
CategoryComputer Hardware
LanguageEnglish

Related product manuals