Condensed Operating Instructions (Continued)
Table COI-1: Test Parameter Setup
Step Test Parameter AC Hipot DC Hipot IR GC Range
1 To enter
programming mode
[F1] =
PROGRAM
[F1] =
PROGRAM
[F1] =
PROGRAM
[F1] = PROGRAM
2 Select Test Step [F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
1-10
3 Select Test Mode [F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
AC, DC, IR, GC, PA
4 Set Test Voltage
Set Current (GC)
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0.05-5kV AC
0.05-6kV DC
0.05-1kV IR
0.1A GC
5 Set High Limit *
Set Low Limit (IR)
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0.001-20mA AC
0.0001-5mA DC
0.1-50000M IR
0, 0.1-5 GC
6 Set Test Time
Set Dwell (GC) ***
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0, 0.1-999s AC
0, 0.1-999s DC
0, 0.1-999s IR
0.1-1s GC
7 Set Low Limit **
Set High Limit (IR)
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0-20mA AC
0-5mA DC
0-50G IR
0-5 GC
8 Set ARC Limit [F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
1-20mA AC
1-5mA DC
9 Set Ramp Time [F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0-999s AC
0-999s DC
0-999s IR
10 Set Dwell Time*** [F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0-999s DC
0-999s IR
11 Set Fall Time [F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0-999s AC
0-999s DC
0-999s IR
12 Set In-Rush Current [F1] = INC.
[F3] = NEXT
0, 0.5uA-5mA DC
13a To program next test
step OR
[F1] = NEW
OR
[F1] = NEW
OR
[F1] = NEW
OR
[F1] = NEW
OR
Program next step
OR
13b To exit
programming mode
[F4] = EXIT [F4] = EXIT [F4] = EXIT Exit programming
* High limit decision is made throughout test time.
** Unit will make low limit decision at the end of test.
* ** Limits will be checked at end of dwell time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the QuadTech Sentry Plus Series and is the answer not in the manual?

QuadTech Sentry Plus Series Specifications

General IconGeneral
BrandQuadTech
ModelSentry Plus Series
CategoryTest Equipment
LanguageEnglish

Related product manuals