Condensed Operating Instructions (Continued)
Table COI-1: Test Parameter Setup
Step Test Parameter AC Hipot DC Hipot IR GC Range
1 To enter
programming mode
[F1] =
PROGRAM
[F1] =
PROGRAM
[F1] =
PROGRAM
[F1] = PROGRAM
2 Select Test Step [F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
1-10
3 Select Test Mode [F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
[F1] = UP
[F3] = NEXT
AC, DC, IR, GC, PA
4 Set Test Voltage
Set Current (GC)
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0.05-5kV AC
0.05-6kV DC
0.05-1kV IR
0.1A GC
5 Set High Limit *
Set Low Limit (IR)
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0.001-20mA AC
0.0001-5mA DC
0.1-50000MΩ IR
0, 0.1-5Ω GC
6 Set Test Time
Set Dwell (GC) ***
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0, 0.1-999s AC
0, 0.1-999s DC
0, 0.1-999s IR
0.1-1s GC
7 Set Low Limit **
Set High Limit (IR)
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0-20mA AC
0-5mA DC
0-50GΩ IR
0-5Ω GC
8 Set ARC Limit [F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
1-20mA AC
1-5mA DC
9 Set Ramp Time [F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0-999s AC
0-999s DC
0-999s IR
10 Set Dwell Time*** [F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0-999s DC
0-999s IR
11 Set Fall Time [F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
[F1] = INC.
[F3] = NEXT
0-999s AC
0-999s DC
0-999s IR
12 Set In-Rush Current [F1] = INC.
[F3] = NEXT
0, 0.5uA-5mA DC
13a To program next test
step OR
[F1] = NEW
OR
[F1] = NEW
OR
[F1] = NEW
OR
[F1] = NEW
OR
Program next step
OR
13b To exit
programming mode
[F4] = EXIT [F4] = EXIT [F4] = EXIT Exit programming
* High limit decision is made throughout test time.
** Unit will make low limit decision at the end of test.
* ** Limits will be checked at end of dwell time.