Page 40 of 85 Operation
Programming an DC Hipot Test (Continued)
END DC Hipot Test Programming. After selecting the I-RUSH limit, press [F1] = NEW to change the step
number in the highlighted box (1-10) and start programming STEP 4-10, OR
press [F4] = EXIT to exit
programming function and return to STAND BY status.
For this example, press [F1] = NEW and proceed to program step 4 as an IR Test (ΒΆ 2.7).
NOTE: The In-Rush current range is dependent on the programmed High Limit. Refer to Safety Features in the
Specifications section, page 8.
Set
Arc Limit
Set arc current limit:
1-5mA in 0.1mA increments
To move highlighted box to RAMP.
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK1 - 5mA 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOWDCSTEP 3/3
VOLT
HIGH
TIME
0.015mA
OFF
OFF
OFF:
:
:
:
:
:
:
2.500kV
2.999mA
5.0s
:I-RUS OFF
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK0 - 999s 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOWDCSTEP 3/3
VOLT
HIGH
TIME
To move highlighted box to I-RUS.
To set Fall Time:
0 - 999s in 0.1s increments
Set
Fall Time
NEW
MORE..
EXIT
NEXT
RMT ERROFSTLOCK1 - 10
F1
F4
F3
F2
FALL
RAMP
ARC
LOWDCSTEP
VOLT
HIGH
TIME
2.500kV
To set next test step = 4
OFF
3.0s
OFF
0.015mA
3.0s
3.0s
OFF
0.015mA
To
exit
programming mode and return
to STAND BY status.
:
:
:
:
:
:
:
5.0s
2.500kV
2.999mA
:
:
:
:
:
:
:
5.0s
2.999mA
3/3
Set
Next
Test Step
[F1]
: Program Step 4
or
[F4] : Exit Program Mode
Set
I-Rus Limit
Set in-rush current limit:
0, 0.5uA -5mA
in 0.001mA increments
To move highlighted box to STEP.
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK0, 0.5uA-5mA 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOWDCSTEP 3/3
VOLT
HIGH
TIME
0.015mA
OFF
3.0s
3.0s:
:
:
:
:
:
:
2.500kV
2.999mA
5.0s
I-RUS
:
:I-RUS OFF
OFF
I-RUS : OFF
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK0 - 999s 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOWDCSTEP 3/3
VOLT
HIGH
TIME
To set Ramp Time:
0 - 999s in 0.1s increments
To move highlighted box to DWELL.
Set
Ramp Time
OFF
OFF
OFF
0.015mA
:
:
:
:
:
:
:
2.500kV
2.999mA
5.0s
OFFI-RUS
:
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK0 - 999s 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOWDCSTEP 3/3
VOLT
HIGH
TIME
To set Dwell Time:
0 - 999s in 0.1s increments
To move highlighted box to FALL.
Set
Dwell Time
OFF
OFF
OFF
0.015mA
:
:
:
:
:
:
:
2.500kV
2.999mA
5.0s
OFFI-RUS :
DWELL :
:
OFF
DWELL
DWELL
:
OFF
3.0s
:
:
DWELL 3.0s
DWELL 3.0s
3.0sDWELL
: