User Manual Ultra
Version M | # DOC00040 April 22, 2012

Questions and Answers

  • D
    Destiny ChapmanSep 23, 2025
    What does voltage over setting mean on Quantel Ultra Big Sky Laser Measuring Instruments?
    • C
      chase34Sep 23, 2025
      This indicates a problem with the charging system, which is a potentially serious safety issue. Contact Quantel Customer Service.
  • S
    Samantha MurphyAug 25, 2025
    What to do if Quantel Ultra Big Sky Laser Measuring Instruments show low coolant level?
    • C
      Cory JonesAug 25, 2025
      If the Quantel Measuring Instruments system is low on coolant, ensure it is properly filled after draining. Refer to the instructions in Maintenance on page 68.
  • K
    Kathryn HowardAug 25, 2025
    What to do if Quantel Measuring Instruments display charger/simmer over temp?
    • B
      Barbara DavisAug 26, 2025
      If the charger inductor is too hot, turn OFF power to the ICE450 and allow it to cool. Make sure that airflow is not obstructed at the front or back of the ICE450. If the interlock condition persists, there may be a problem with the charger circuitry. In this case, contact Quantel Customer Service.
  • L
    Laura WilsonAug 27, 2025
    Why is the coolant temperature high in my Quantel Measuring Instruments?
    • D
      Donna DavisAug 27, 2025
      If the coolant is too warm for proper operation, turn off the ICE450 and allow it to cool. The high-temperature setting may need to be increased, or the cooling fans may have failed. Contact Quantel Customer Service for these issues.
  • A
    Amanda TerrellAug 27, 2025
    What causes low coolant temperature in Quantel Ultra Big Sky Laser Measuring Instruments?
    • E
      Emily BryantAug 27, 2025
      If the coolant is not warm enough for proper operation, allow time for the coolant to warm. The low-temperature setting may need to be reduced, or the heater minimum power setting may need adjustment. The heater circuit may have failed. Contact Quantel Customer Service for these issues.
  • A
    April MerrittSep 14, 2025
    Why is there no system power in my Quantel Ultra Big Sky Laser?
    • J
      James AdamsSep 14, 2025
      If there is no power connected, check the simple things first—make sure the power cord is plugged into the outlet and making connection to the back of the ICE.
  • R
    Robert LeeSep 12, 2025
    How to resolve flashlamp disabled time out delay expired on Quantel Ultra Big Sky Laser Measuring Instruments?
    • J
      Jason OrtegaSep 12, 2025
      If the flashlamp timeout has expired, disable timeouts or change the flashlamp timeout to 00:00.
  • M
    Michael AshleySep 13, 2025
    What to do if simmer disabled time out delay expired on Quantel Measuring Instruments?
    • K
      kelly73Sep 13, 2025
      If the simmer timeout has expired, you should not attempt to repair this problem. Contact Quantel Customer Service for repair.
  • S
    Sean TurnerSep 4, 2025
    What to do if Quantel Ultra Big Sky Laser Measuring Instruments show defective temp. sensor error?
    • L
      Leslie JamesSep 5, 2025
      If one or more of the three feedback thermistor circuits (coolant, charger/ simmer, and harmonic generator) is faulty or is running too cold, ensure all temperatures are above 18 °C. If the system does not include a harmonic generator, make sure the HG option is turned off from the System Info menu.
  • K
    Kendra HartSep 2, 2025
    How to fix power too high error in Quantel Ultra Big Sky Laser?
    • J
      James SmallSep 2, 2025
      If the voltage or frequency settings are too high, resulting in power too high, reduce voltage, energy, or frequency until the power is less than 450W. Verify this and save the configuration.

Summary

LASER SAFETY

Hazard Information

Defines terms and instructions for care, use, and handling of Class IV laser equipment.

Terms & Warning Symbols

Explains hazard symbols (Danger, Warning, Caution, Note) and their meanings.

General Hazards

Describes general hazards and unsafe practices causing damage or injury.

Other Hazards

Lists typical hazards such as exposure, voltage, and lifting risks.

Safe Operation of the Laser

Details essential precautions for safe laser operation.

SAFETY LABELS

SETUP

Connecting System Components

Details for connecting the ICE450, Laser Head, and Remote Box for upright and rack systems.

Filling the Standard ICE450

Procedure for filling the coolant reservoir of the standard ICE450 unit.

Filling the Rack ICE450

Procedure for filling the coolant reservoir of the rack-mounted ICE450 unit.

FUNCTIONS

OPERATION

Safety Precautions

General safety guidelines and precautions for operating the laser system.

Precautions for Laser Operation

Specific guidelines to prevent laser damage and ensure proper operation.

Remote Interlock System

Explains the function and connection of the remote interlock for safety shutdown.

Remote Box Controls and Indicators

Details the controls and indicators on the Remote Box unit.

Automatic Mode (INT;INT)

Explains configuration and operation in automatic mode.

External Modes (EXT)

Discusses external triggering requirements and methods.

Example Start-Up Procedure

Provides a step-by-step guide to starting the laser system.

Shutdown Procedure

Details the steps for safely shutting down the laser system.

Decreasing Output Energy

Explains how to decrease and increase laser output energy.

Increasing Output Energy

Details methods for increasing laser output energy.

SOFTWARE

TECHNICAL SPECIFICATIONS

MAINTENANCE

TROUBLESHOOTING

Resolving Interlock Fault Conditions

Details common interlock values, messages, causes, and solutions.

Diagnosing Problems

General guide for identifying and resolving system issues.

Quantel Ultra Big Sky Laser Series Specifications

General IconGeneral
BrandQuantel
ModelUltra Big Sky Laser Series
CategoryMeasuring Instruments
LanguageEnglish