What should I do if the Laser Head housing is opened on Quantel Ultra ICE 450 Measuring Instruments?
- AAngela JohnsonSep 6, 2025
If the I/O cable is not connected after opening the Laser Head housing, ensure the I/O cable is fully seated.
What should I do if the Laser Head housing is opened on Quantel Ultra ICE 450 Measuring Instruments?
If the I/O cable is not connected after opening the Laser Head housing, ensure the I/O cable is fully seated.
What to do if Quantel Ultra ICE 450 Measuring Instruments show low coolant level?
If the Quantel Measuring Instruments system is low on coolant, ensure it is properly filled after draining. Refer to the instructions in Maintenance on page 68.
What to do if Quantel Ultra ICE 450 Measuring Instruments display charger/simmer over temp?
If the charger inductor is too hot, turn OFF power to the ICE450 and allow it to cool. Make sure that airflow is not obstructed at the front or back of the ICE450. If the interlock condition persists, there may be a problem with the charger circuitry. In this case, contact Quantel Customer Service.
Why is the coolant temperature high in my Quantel Ultra ICE 450 Measuring Instruments?
If the coolant is too warm for proper operation, turn off the ICE450 and allow it to cool. The high-temperature setting may need to be increased, or the cooling fans may have failed. Contact Quantel Customer Service for these issues.
What causes low coolant temperature in Quantel Ultra ICE 450 Measuring Instruments?
If the coolant is not warm enough for proper operation, allow time for the coolant to warm. The low-temperature setting may need to be reduced, or the heater minimum power setting may need adjustment. The heater circuit may have failed. Contact Quantel Customer Service for these issues.
Why is there no system power in my Quantel Ultra ICE 450?
If there is no power connected, check the simple things first—make sure the power cord is plugged into the outlet and making connection to the back of the ICE.
How to resolve flashlamp disabled time out delay expired on Quantel Measuring Instruments?
If the flashlamp timeout has expired, disable timeouts or change the flashlamp timeout to 00:00.
What to do if simmer disabled time out delay expired on Quantel Ultra ICE 450 Measuring Instruments?
If the simmer timeout has expired, you should not attempt to repair this problem. Contact Quantel Customer Service for repair.
What to do if Quantel Ultra ICE 450 Measuring Instruments show defective temp. sensor error?
If one or more of the three feedback thermistor circuits (coolant, charger/ simmer, and harmonic generator) is faulty or is running too cold, ensure all temperatures are above 18 °C. If the system does not include a harmonic generator, make sure the HG option is turned off from the System Info menu.
How to fix power too high error in Quantel Measuring Instruments?
If the voltage or frequency settings are too high, resulting in power too high, reduce voltage, energy, or frequency until the power is less than 450W. Verify this and save the configuration.
Category | Measuring Instruments |
---|---|
Frame Grabber | Yes |
Display | LCD |
Weight | 2.5 kg |
Power Supply | 100-240 VAC, 50/60 Hz |
Dimensions | 300 x 200 x 100 mm |
Explains hazard terms, symbols, and instructions for safe laser use.
Details hazard symbols and their meanings for safe operation.
Details general hazards and unsafe practices associated with laser use.
Lists typical hazards associated with the product and its handling.
Outlines essential safety procedures for operating the laser system correctly.
Instructions for securing the Laser Head to a flat mounting surface.
Step-by-step guide for connecting the standard upright Ultra System components.
Instructions for connecting the components of a rack-mounted Ultra System.
Detailed procedure for filling the coolant reservoir of the standard ICE450 unit.
Detailed procedure for filling the coolant reservoir of the rack-mounted ICE450 unit.
Describes the functions of the ICE450 front panel controls and indicators.
General safety precautions and guidelines to prevent system damage.
Details the Remote Box buttons, navigation, and basic menu structure.
Provides detailed explanations of the Remote Box menu items.
Step-by-step guide for starting the laser system.
Procedure for safely shutting down the laser system.
Information on the RS-232 serial interface for system control and communication.
Step-by-step guide for completely draining the ICE450 coolant.
Procedure for completely draining the coolant from a rack-mounted ICE450.
Procedure for replacing the flashlamp for optimal performance.
Provides solutions for common interlock fault conditions and their causes.
Troubleshooting steps for serial communication issues with the ICE450.
Guide to diagnosing general system problems and their potential causes.