MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
9
Door Open detected
Close door and press RETEST
(Press IGNORE only if faulty sensor)
IGNORE RETEST
* Open and Close the door
* Press IGNORE if operator
feels
door is securely even when
alarm is repeating itself
* Contact service
immediately
10, 11
System Config mismatch Error
Call Service
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
12
Left Pump failed diagnostics
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
13
Right pump failed diagnostics
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
14
Left pump failed diagnostics
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
15
Right pump failed diagnostics
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
16
Pneumatic leak or compressor
malfunction
Consider attaching medical air if
available
Consult Service after this case
SHUTDOWN RESTART
* Connect medical air
* Contact service following
procedure
17
Mechanism Valve failed diagnostics
RESTART CONFIRM
* Open door and manually
depress mechanism valves
several times
* Restart System
* Call service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action
5

Other manuals for Quest Engineering MPS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quest Engineering MPS 3 and is the answer not in the manual?

Quest Engineering MPS 3 Specifications

General IconGeneral
BrandQuest Engineering
ModelMPS 3
CategoryMedical Equipment
LanguageEnglish