MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
Door Open detected
Close door and press RETEST
(Press IGNORE only if faulty sensor)
IGNORE RETEST
* Open and Close the door
* Press IGNORE if operator
feels
door is securely even when
alarm is repeating itself
* Contact service
immediately
System Config mismatch Error
Call Service
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
Left Pump failed diagnostics
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
Right pump failed diagnostics
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
Left pump failed diagnostics
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
Right pump failed diagnostics
RESTART CONFIRM
* Restart System
* Call service after multiple
occurrences
Pneumatic leak or compressor
malfunction
Consider attaching medical air if
available
Consult Service after this case
SHUTDOWN RESTART
* Connect medical air
* Contact service following
procedure
Mechanism Valve failed diagnostics
RESTART CONFIRM
* Open door and manually
depress mechanism valves
several times
* Restart System
* Call service after multiple
occurrences