MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
4.3. Alarm Code List and Solutions
The following table lists alarm (error) codes, the displayed message, and
recommended corrective actions. The alarm code can be seen just below the
Caution and Warning symbol in the message window. A chronological listing of all
alarm codes on the device can be access by selecting the Error Report button
form the Menu screen.
Alar
m
Cod
e
Message Display
Corrective Action
1
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
2
System Data Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
3, 4
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
5
Door is open
Close the door to continue
SHUTDOWN
* Close the door
* Press shutdown to turn
System off
6
Door is open
Close door to continue
* Close the door
7
Door Open still detected
Press IGNORE only if faulty sensor
IGNORE RETEST
* Open and Close door
* Press IGNORE if operator
feels door is securely closed
but
the console is not recognizing
that
* Contact service following
procedure
8
Door Open detected
Open then Close door
* Open and Close the door
* Contact service following
procedure
4

Other manuals for Quest Engineering MPS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quest Engineering MPS 3 and is the answer not in the manual?

Quest Engineering MPS 3 Specifications

General IconGeneral
BrandQuest Engineering
ModelMPS 3
CategoryMedical Equipment
LanguageEnglish