MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
142
Pneumatic leak or internal compressor
malfunction
Consider attaching medical air if
available
Consult service after this case
CONFIRM
* Connect to medical air
* Call service following
procedure
143, 144
Compressor malfunction detected
Consult service after this case
CONFIRM
* Connect to medical air
* Call service following
procedure
145
Pneumatic system failure
Connect medical air
Service required
SHUTDOWN CONFIRM
* Connect to medical air
* Call service following
procedure
146
System Communication Error
Manual Restart Required
* Manually Restart System
* Call service after multiple
occurrences
147
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
148
System Communication Error
Manual Restart Required
* Manually Restart System
* Call service after multiple
occurrences
149
Temperature sensor Fail
Call Service after this Case
CONFIRM
* Safe to use. Board
temperature
sensor is not essential for
operation
* Call service after multiple
occurrences
Alar
m
Cod
e
Message Display
Corrective Action