MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
Drug pump LED fail
Contact Service after this Case
CONFIRM
* Safe to use. Drug pump is
fully
functional. Only the LED may
be compromised.
* Call service after multiple
occurrences
Board Temperature sensor Fail
Call Service after this Case
CONFIRM
* Safe to use. Board
temperature
sensor is not essential for
operation.
* Call service after multiple
occurrences
Elevated internal temperature detected
Check for obstructed air flow
Maintenance required
CONFIRM
* Check fans and air inlets for
obstructions
* Restart System
* Contact service after multiple
occurrences
Internal temperature Error detected
Check for obstructed air flow
Unable to proceed. Call for Service
SHUTDOWN
* Check fans and air inlets for
obstructions
* Restart System
* Contact service after multiple
occurrences
System Communication Error
Manual Restart Required
* Manually Restart System
* Call service after multiple
occurrences