MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
Battery temperature elevated
Turn off switch on the rear panel
Unable to proceed. Call for service
* Check fans and air inlets for
obstructions
* Restart System
* Contact service after multiple
occurrences
Battery Charge Error
Battery backup unavailable
Consult service after this case
CONFIRM
* Call service after multiple
occurrences
Battery Charge Error
Consult service after this case
CONFIRM
* Call service after multiple
occurrences
Air detected in delivery Line
IGNORE CONFIRM
* Check delivery line to ensure
it
is fully inserted in sensor
* If sensor is ignored, manually
inspect delivery line for air
moving forward
Air detected in delivery line. Flow
continues
Manually check for air
CONFIRM
* Check delivery line to ensure
it
is fully inserted in sensor
* If sensor is ignored, manually
inspect delivery line for air
moving forward
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences