MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
225
Additive pump error
Additive pump is Off
CONFIRM
* Check Additive Cartridge to
verify it is properly installed
and
fully secured to the pump
* If the cartridge was secure, try
removing cartridge and
reinserting it
* If the cartridge was secure, try
stopping and restarting
cardioplegia flow
* If error still persists, restart
console and try again
* Contact service after multiple
occurrences
226
Internal H2O temperature sensor error
Ensure proper Cold water source
Call Service after this Case
CONFIRM
* Continue as normal
OR
* Restart and resume case
* Call service after multiple
occurrences
228
Battery Error
Consult service after this case
CONFIRM
* Continue as normal
OR
* Restart and resume case
* Call service after multiple
occurrences
229, 230,
231, 232
Board Temperature sensor Fail
Call Service after this Case
CONFIRM
* Continue as normal. Board
temperature sensor is not
essential for operation.
* Call service after multiple
occurrences
235
System Data Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
alarms
Alar
m
Cod
e
Message Display
Corrective Action
35

Other manuals for Quest Engineering MPS 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quest Engineering MPS 3 and is the answer not in the manual?

Quest Engineering MPS 3 Specifications

General IconGeneral
BrandQuest Engineering
ModelMPS 3
CategoryMedical Equipment
LanguageEnglish