10A96 Ion chamber Adapter
This adapter adapts the 10X5 ion chambers to the 9660 ion
chamber digitizer. It contains a microprocessor that translates
between the jumpers, resistors and potentiometer located within
the probe stem of 10X5-series ion chambers into signals that
mimic the EAROM that holds calibration and chamber-type
information the Accu-Pro™ requires. A temperature sensor is
included to provide that information also. Its diameter is slightly
larger than that of the chamber stem, with a hard-mounted 1015
connector on one end and a 9096-style connector on the
opposite end. It draws power from the 9660 Ion Chamber
digitizer.