! Dent(al) thr(eshold) SELECT
Raised threshold for kV-Dental. Values are multipliers of the kV pulse threshold. 1, 10, 25, 50,
100, 250
, 600, and 1500.
! Pulsed Rad Mod Index no. SELECT
Set up presets 0 thru 5 for Pulsed/Cine mode. For each, select four parameters:
Pulse Frequency
00.25.512345
66.2577.5891012.5
15 25 30 40 50 60 75 90
Trig threshold.
Number is relative dose rate. () is dose rate in nR/sec for 10X6-6
3 (55) 5(92) 10 (184) 18 (330) 30 (550)
55 (1000) 100 (1840) 182 (3300) 300 (5500) 500 (9200)
Acquire Delay
Number of frames: 1 - 256
converted to time based on selected pulse frequency
Acquire Interval
Number of frames: 1 - 256
converted to time based on selected pulse frequency
! Press SELECT to exit Setup
If any changes have been made, you then will see
Accept Changes?
> OK
select accepts the changes and returns to main menu.
UP or DN changes OK to Cancel and then select returns to the main menu discarding the changes.