90M10 mAs Sensor
Battery-powered, isolated, clamp-on non, invasive mA sensor
Aperature 23mm.
Synchronized to Accu-kV measurement
Real-time waveform.
Automatic zero and no range switching.
2000 mA or 9999 mAs full-scale
Measures absolute value of mA.
Resolution greater of 3-4 digits or .015mA
mA accuracy:
±4% of reading (Limited by 0.6mA RMS noise below 15mA).
mAs accuracy (1-s pulse):
±4% of reading (Limited by 0.6mAs RMS noise below 15ma).
77 μs sample rate.
-2.6% at 720Hz, -3 dB at 2.33 kHz.
Digital waveform sent over the USB port to XLPRO
Automatic turn-on only when kV function is active
Battery lifetime: Approximately 35 operating hours.
Caution: In the presence of strong electromagnetic fields, performance may degrade up to 1 Amp.