EasyManuals Logo

Radcal Accu-Pro User Manual

Default Icon
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
40X M Sensors:
Potential 22 kV - 40 kV
Accuracy 0.5 kV for Mo/Mo
±1 kV for Hologic Selenia x-ray machines with W anode and Rh filter
Repeatability (COV of 10 150 ms exposures)
0.2kV for Mo/Mo
0.1kV for Hologic Selenia x-ray machines with W anode and Rh filter
Rotation (normal to source-detector line)
about sensor long axis: 0.5 kV for 6º
about sensor short axis: 0.5 kV for 5º
Alignment (tube anode-cathode vs sensor long axis)
0.5 kV for 10º
Long-term drift: 0.5 kV for 1 yr
Width accuracy (2 ms - 5 sec): 0.1% of value + 0.2 ms
Ripple Error (Displayed kVp): 0.5 kV for 10%, 0-to-peak, ripple < 4 kHz
Filtration
25 - 35 μm Mo 0.1kV/μm Mo
28 - 40 kV 0.8 kV/mm Al
Reference conditions:
Temperature 21 - 23 C
Mo Anode, 30 μm Mo total filtration
Anode current 50 mA, <1% ripple
Focus - sensor distance 30 cm
Sensor long axis perpendicular to anode-cathode axis
Sensor plane normal to focus-sensor line
Width measured at 75% of kV waveform
24

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Radcal Accu-Pro and is the answer not in the manual?

Radcal Accu-Pro Specifications

General IconGeneral
BrandRadcal
ModelAccu-Pro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals