EasyManuals Logo

Radcal Accu-Pro User Manual

Default Icon
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
21
54
12Ø
15Ø
16
38
12
2 MeV
Build-up
Cap
Center of
Sensitive
Volume
Dimensions in millimeters
10X6-0.6
Exposure Rate 20 μR/s - 133
R/s
200 nGy/s - 1.17
Gy/s
Dose 100 μR - 589
kR
1 μGy - 5 kGy
Auto Dose
Threshold
22 mR/s 189 μGy/s
Calibration 4% @
60
Co
Rate
Dependence
2%, 10 mR/s - 100 R/s
Energy
Dependence
5%, 40 keV - 1.33 Mev
(with buildup cap)
Construction C552 air-equivalent walls &
electrode, 0.6 cm
3
active volume.
12 m cable
Application High-intensity gamma radiation in-
beam measurements. Irradiators.
27

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Radcal Accu-Pro and is the answer not in the manual?

Radcal Accu-Pro Specifications

General IconGeneral
BrandRadcal
ModelAccu-Pro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals