- 41 - 
 
Deleting tare 
Entered tare value can be deleted by pressing 
 button on the overlay or by entering tare value of 
0.000g (see description above). 
 
11.2.4.  WEIGHING Mode Settings 
The software allows setup of operating parameters (filters, value release and autozero function, 
deleting the last digit and other settings) separately for each working mode. 
It enables customizing the instrument and utilizing its properties depending on the user’s needs and 
expectations, or on specific requirements for selected working mode (e.g. DOSING); as a result the 
device operation is quick and easy. 
 
Filter level setting 
Filter settings adjustment depends on the working environment. For the best possible conditions the 
filter can work in a very fast mode (V.FAST value for Filter parameter); however, if the conditions are 
poor (shakes, drafts), the filter should be set to slow or very slow option (SLOW or V. SLOW value for 
Filter parameter). The effectiveness of the filter is different throughout the weighing range. The filter 
works slower when “approaching” the weighed mass, it works more intensively for weighed mass 
within the set range of the filter (the parameter for setting filter range is accessible only from the 
service menu – the user does not have any access to it).  
Depending on the filter, the weighing time is shorter (V.FAST and FAST) or longer (SLOW and V. 
SLOW).  
 
 
Value release 
Since ambient conditions at a workplace vary, it is necessary to determine the value release 
parameter in the most preferable way enabling balance adaptation, parameter options are: 
FAST.+REL., FAST or RELIABLE. Depending on the selected option, weighing time is either shorter 
or longer. 
 
 
Autozero function 
The software features an autozero function (Auto) ensuring precise mass indication. This function 
automatically controls and corrects zero indication. When Autozero is enabled, it compares balance 
indications at declared time intervals e.g. 1s, provided that weighing pan is unloaded and display 
indication is close to zero. If results vary less than the declared AUTOZERO range e.g. one division, 
balance zeroes automatically, marker of stable measurement result – 
, and precise zero marker – 
 are displayed.