- 90 -
An example:
Command:
US_mg CR LF
set „mg” unit
Response:
US_mg_OK CR LF
„mg” set as a current unit
Give current unit
Command overview: command returns current unit.
Format: UG CR LF
Response options:
UG_x_OK CR LF
command carried out, response: current unit
UG_I CR LF
command understood but not accessible at this moment
x parameter, unit symbol
An example:
Command:
UG CR LF
return current unit
Response:
UG_ct_OK CR LF
currently set unit is „ct”
Activate sound signal
Command overview: command activates BEEP sound signal for a specified amount of time
Format: BP_CZAS CR LF
Response options:
BP_OK CR LF
command carried out, BEEP sound signal activated
BP_E” CR LF
no parameter or incorrect format
BP_I CR LF
command understood but not accessible at this moment
TIME parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 ÷ 5000>.
If value greater than the permissible high limit is given, than BEEP sound is operated for the
maximum permissible amount of time.
An example:
Command:
BP_350 CR LF
activate BEEP for 350 ms
Response:
BP_OK CR LF
BEEP activated
CAUTION!
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is activated by
means of other device: keypad, touch panel, proximity sensors.
Send all implemented commands
Format: PC CR LF
Command:
PC CR LF
- send all implemented commands
Response:
PC_A_”Z,T,S,SI…”
- command carried out, the terminal displays all
implemented commands.
Give balance type
Format: BN CR LF
Response options:
BN_A_”x” CR LF
- command understood, response: balance type
BN_I CR LF
- command understood but not accessible at this moment
x series of types for a particular balance (in between inverted commas), with general balance type
in front
An example:
Command:
BN CR LF
return balance type
Response:
BN_A_”AS”
balance type: ”AS R”
Give max capacity
Format: FS CR LF
Response options:
FS_A_”x” CR LF
- command understood, response: Max capacity
FS_I CR LF
- command understood but not accessible at this moment
xMax value of reading units (in between inverted commas)
An example:
Command:
FS CR LF
return Max capacity
Response:
FS_A_”220.0000”
Max capacity: ”220 g”
Give program version
Format: RV CR LF