Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
832User Manual 1178.8520.02 ─ 01
Suffix: 
<ri>
.
1..n
Selects the measurement range.
Parameters:
<SweepTime> Sweep time.
The range depends on the ratios of the span to the RBW and
the RBW to the VBW. Refer to the data sheet for more informa-
tion.
Manual operation:  See " Sweep Time " on page 257
[SENSe:]LIST:RANGe<ri>:SWEep:TIME:AUTO 
<State>
This command turns automatic selection of the sweep time for a spurious emission
measurement range on and off.
Suffix: 
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State> ON | OFF | 0 | 1
*RST:  1
Manual operation:  See " Sweep Time Mode " on page 257
[SENSe:]LIST:RANGe<ri>:TRANsducer <Transducer>
This command selects a transducer factor for a spurious emission measurement
range.
Note that
●
the transducer must cover at least the span of the range
●
the x-axis has to be linear
●
the unit has to be dB
Suffix: 
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Transducer> String containing the transducer file name, including the path
information.
Manual operation:  See " Transducer " on page 258
14.5.7.3  Configuring the List Evaluation
The following commands configure the list evaluation.
Configuring and Performing Measurements