M3SR Series 4100 Synthesizer Calibration
7.10 Operating Manual 6175.4760.02 – 01
7.1.1.3 Possible Errors during Calibration
To get the specific error of a faulty calibration process it is necessary to enter the error list of 
the radio.
The following 3 errors are related to the calibration process.
LOW RXREF LEVEL:
This error message indicates that the output power of the frequency standard is too low. 
Please increase the output power of the frequency standard or use a frequency standard that 
meets the required specification.
WRONG REF FREQ:
This error message indicates that it was not possible to perform a calibration with the fre-
quency form of the frequency standard. It should be checked if the output frequency of the 
frequency standard is set as Rx frequency at the radio. Please retry the calibration; if the er-
ror appears again, the radio or the frequency standard is defective. Under this condition 
please contact the service.
REFTUNE LIMIT:
This error message indicates that the limit of the tuning range was reached and it was not 
possible to find valid calibration values. Provided that the output frequency of the frequency 
standard is set as Rx frequency at the radio and the frequency standard meets the required 
specification, the radio may be defective. In this case please contact the service.